The KT-M high accuracy temperature converter converts signals from thermocouples or resistance thermometer into high accuracy digital signals, and outputs the digital signals to a recorder or PC.
Specifications | Models | ||
---|---|---|---|
Main unit (base unit) | 24 channel | KT-M301 | |
6 channel | KT-M311 | ||
Input unit | Thermocouple Input | N Thermocouple | KT-M411N |
B thermocouple | KT-M411B | ||
S thermocouple | KT-M411S | ||
R thermocouple | KT-M411R | ||
K thermocouple | KT-M411K | ||
E thermocouple | KT-M411E | ||
J Thermocouple | KT-M411J | ||
T thermocouple | KT-M411T | ||
resistance thermometer input | Pt100Ω resistance thermometer | KT-M411P | |
Pt-Co resistance thermometer | KT-M411C |
Models | KT-M301 | KT-M311 |
---|---|---|
Input Type | Thermocouple: K, E, J, T, B, R, S, N resistance thermometer: Pt100 Pt-Co 4-wire, 3-wire |
|
Number of input points | Thermocouples: up to 24 points resistance thermometer: up to 12 points Thermocouples and resistance thermometer can be mixed. However, one resistance thermometer will occupy the space of two thermocouples. |
Thermocouple: Maximum 6 points (Maximum 24 channel with 4-unit multi-drop) resistance thermometer: Maximum 3 points (Maximum 12 channel with 4-unit multi-drop) Thermocouples and resistance thermometer can be mixed. However, one resistance thermometer will occupy the space of two thermocouples. |
communication interface | RS-485 (MODBUS protocol) Capture cycle: Maximum 1 second/24 channels (1:1 connection only) |
|
Power supply | 100-240VAC 50/60Hz ±2% | 100-240VAC 50/60Hz ±2% |
Power consumption | 約50VA | 約15VA |
External dimensions | W 500×D 200×H 120mm | W 320×D 151.5 ×H 115mm |
Weight | Approx. 4.3kg (maximum) | Approx. 1.9kg (with 6ch installed) |
Accuracy guaranteed range | Temperature: 18 to 28°C, Humidity: 45 to 65% RH (no dew condensation) | Temperature: 0 to 50°C, Humidity: 45 to 65% RH (no dew condensation) |
Scope of use | Temperature: 0 to 40°C / Humidity: 20 to 80% RH (no dew condensation) | Temperature: 0 to 50°C / Humidity: 20 to 80% RH (no dew condensation) |
The KT-M is a high accuracy temperature converter that converts signals from thermocouples or resistance thermometer into high accuracy digital signals and outputs the digital signals to a recorder or computer.
KT-M
Please use it for traceable temperature and humidity measurement and maintaining the accuracy of sensors.
If you register as a CHINO Web member, you will be able to view the product's instruction manual.
If you wish, we will send you the latest information on CHINO products via email.
You must be a login to download.