Frequently Asked Questions (FAQ)

There are class 1 and class 2 K thermocouples, and looking at the JIS standard table, class 1 is listed as up to 1000°C, and class 2 is listed as up to 1200°C. Are the measurable temperatures different depending on the class?
Class 1 and Class 2 refer to different levels of accuracy, not the range of temperatures that can be measured.
The temperature range guaranteed by JIS differs depending on the class. Class 1, 1000°C and above, is outside the JIS standard, so accuracy is not guaranteed, but temperature measurement is possible.

If you have any problems with temperature control,
 please feel free to contact us.

Membership
CHINO Web Membership Information

If you register as a CHINO Web member, you will be able to view the product's instruction manual.
If you wish, we will send you the latest information on CHINO products via email.

You must be a login to download.