Intrinsically safe explosion proof platinum/cobalt sheathed resistance thermometer NR800-6, NR800-7 (Japan Stadnards)

Intrinsically safe explosion proof type platinum cobalt sheath resistance thermometer NR800-6 and NR800-7 are for cryogenic use resistance thermometer of the intrinsically safe explosion proof type. insulation Combined with barriers, they have passed the certification for intrinsically safe explosion proof equipment by the Japan Industrial Safety Technology Association (designate type verification agency of the Ministry of Health, Labor and Welfare).

Passed the intrinsic safety explosion proof equipment inspection by the Technology Association of Industrial Safety when combined with a barrier

  • Capable of measuring from 4K to 375K with an accuracy of ±0.5K
  • Improved insulation and robust by using φ1.6 sheathed wire between the temperature sensing part at the tip of the sensor and the sleeve
  • The sheath wire can be installed in a bent state.
  • When combined with a barrier, it can be used in Class 0 hazardous locations with liquid hydrogen.

Specifications

Sensors Models NR800-6, NR800-7
Measurement temperature NR800-6: 4K to 375K
NR800-7: 15K to 375K
resistance at 0℃ 100Ω±0.15Ω
tolerance ±0.5K
Measurement current 1mA
sensitivity Min 0.09Ω/K (at 12K), max 0.40Ω/K
Protective tube Protective tube: capsule-type oxygen-free copper with gold plating
φ2×30mm
Sheath: SUS316 φ1.6mm
Protective tube + sheath length: 50 to 100,000 mm
Lead 4-wire vinyl-coated copper conductor
Length: 500-600,000mm
explosion proof specifications explosion proof structure types Intrinsically safe explosion proof proof structure
Explosion class and ignition level of the target gas or vapor IICT4 Ga
Permissible voltage of intrinsically safe circuit 12.4V
Permissible current of intrinsically safe circuit 42mA
Intrinsically safe circuit allowable power 69,3mW
Type inspection approval number No. TC22088X

Download Documents & Softwares

Catalog

Specification Sheet (PS Sheet)

CAD Data

*Downloading materials is a members-only service.
*If you wish to download, please login as a member and click "Download."

Instruction Manual

*Downloading materials is a members-only service.
*If you wish to download, please login as a member and click "Download."

Softwares

Application Examples

RelatedInfo

movie

FAQ

If you have any problems with temperature control,
 please feel free to contact us.

Membership
CHINO Web Membership Information

If you register as a CHINO Web member, you will be able to view the product's instruction manual.
If you wish, we will send you the latest information on CHINO products via email.

You must be a login to download.