The high temperature hygrometer HN-Z series uses a zirconia limiting current humidity sensor development in-house, and is capable of in-line measurement of water vapor pressure in high temperature environments above 100°C, up to a maximum of 250°C. By simultaneously measuring oxygen concentration, the effects of changes in oxygen concentration can be corrected even in environments where oxygen concentration changes due to combustion, etc., allowing for more accurate water vapor pressure measurement.
Protection tube length | Models | Model (Installation Method) | Mounting bracket |
---|---|---|---|
400mm | HN-ZS040FC | Fixed flange type | JIS 5K 25R.F |
HN-ZS040LC | Sliding flange type | JIS 5K 25R.F | |
HN-ZS040NC | Fixed nipple type | R1 | |
HN-ZS040SC | Sliding nipple type | R1 | |
900mm | HN-ZS090FC | Fixed flange type | JIS 5K 25R.F |
HN-ZS090LC | Sliding flange type | JIS 5K 25R.F | |
HN-ZS090NC | Fixed nipple type | R1 | |
HN-ZS090SC | Sliding nipple type | R1 |
Measurement range | Models | |
---|---|---|
Output signal: 4 to 20mADC | Output signal: 0 to 10mVDC | |
0 to 10kPa | HN-ZC11C | HN-ZC12C |
0 to 20kPa | HN-ZC21C | HN-ZC22C |
0 to 30kPa | HN-ZC31C | HN-ZC32C |
0 to 50kPa | HN-ZC51C | HN-ZC52C |
0 to 100 kPa | HN-ZCX1N | HN-ZCX2N |
Models | Specifications |
---|---|
HN-ZW05 | 8-core, single-layer shield Cable length: Standard 5m (maximum 30m) |
element | Zirconia Solid Electrolyte |
---|---|
Measurement range | 0 to 10kPa, 0 to 20kPa, 0 to 30kPa, 0 to 50kPa or 0 to 100kPa However, 0 to 100 kPa is only available without oxygen compensation function (depending on converter) |
Accuracy ratings | ±0.3kPa when 0 to 10kPa ±0.6kPa when 0 to 20kPa ±0.9kPa when 0 to 30kPa ±1.5kPa when 0 to 50kPa ±3.0kPa when 0 to 100kPa |
Response time | Within 1 minute (90% response) |
Operating temperature limit | 0 to 250°C (excluding 100mm from terminal box) 0 to 100°C (terminal box, cable) |
Connection method | terminal Connection |
Installation Method | Fixed flange, sliding flange, 4 types: fixed nipple, sliding nipple |
Terminal box | Aluminum die casting |
Protective tube | SUS316 |
filter | SUS316 |
Weight | Approx. 2.0kg (fixed flange, protective tube length 400mm) |
Measurement range | 0 to 10kPa (HN-ZC1) 0 to 20kPa (HN-ZC2) 0 to 30kPa (HN-ZC3) 0 to 50kPa (HN-ZC5) 0 to 100kPa (HN-ZCX) |
---|---|
Display | LED 3-digit digital display |
4 to 20mA DC (load resistance within 500Ω) Or 0 to 10mV DC (output resistance 50Ω) |
|
Operating temperature limit | 0~50℃ |
Power supply | 85-264V AC 50/60Hz |
Power consumption | 最大15VA |
Case | Ordinary Steel Plate |
Installation Method | Wall mounting (2 M5 screws) |
Weight | 約1.2kg |