The MG series oxygen concentration meter uses a limiting current type zirconia oxygen detection element and is equipped with a variety of alarm functions as well as a measurement display. It is a digital oxygen meter that can measure oxygen concentrations from low concentrations as an oxygen deficiency meter or oxygen concentration meter. Wall-mounted types, separate types, measurement monitors, and suction types are available depending on the oxygen concentration measurement application.
Name | Models |
---|---|
Wall-mounted oxygen meter | MG1000-000 |
Oxygen meter with separate detector (2m lead) | MG1220-000 |
Wall Mount Oxygen Sensor | MG2100-000 |
One-point measurement monitor | MG3010-00 |
*The above is for transmission signals of 0 to 250 mV.
Measurement method | Zirconia solid electrolyte system |
---|---|
sampling Method | Natural diffusion type |
Measurement range | 0-25% O2 |
Response time | Within 20 seconds (90% response) |
Accuracy ratings | 0.0~25.0%O 2 …±0.5%O 2 ±1digit 0.00~2.50%O 2 …±0.05%O 2 ±1digit (In a dry gas consisting of oxygen and nitrogen) |
display method | 3-digit yellow-green LED digital display |
Analog Output | 0~250mV DC (output resistance 10Ω) |
Alarm Setpoint | 18% O2 fixation |
Alarm lamp | Yes |
Alarm buzzer | Yes |
Alarm Output | 2 circuits contact capacity: resistance load 110V AC 0.2A, 24V DC 1A |
Operating temperature limit | 0~40℃ |
Operating humidity range | 80% RH or less (but no dew condensation) |
Power supply | 100V AC 50/60Hz |
Power consumption | 最大8VA |
Weight | 約500g |
Mounting Method | Switch box mounting or wall mounting |
Environmental testing rooms, analysis rooms, food fermentation rooms, semiconductor clean rooms, boilers, plant factories
If you register as a CHINO Web member, you will be able to view the product's instruction manual.