Used for setting and display the IM series infrared moisture /thickness gauge.

Features

Supports RS-232C, RS-422A, and RS-485 communication. Can output analog current of 4-20mA.

Specifications

Models IRGMEG3R IRGMEG3A IRGMEG3S
Input Signal RS-485 (connects to detector), up to 9 units can be connected
analog Output analog signal: 4 to 20mADC, 3 outputs
Output 1/2... load resistance 600Ω or less
Output 3... load resistance 400Ω or less
Communication Output RS-232C RS-422A RS-485
Output scaling Numeric keypad settings (0.1 step)
Output Update Period Communication output: 28 ms x number of detector
display ①Measurement value display, □□□□.□ (decimal point position variable)
②Head number, CH.No., parameter display
smoothing T Smoothing time setting for smoothing (delay): T = 0.1 to 99.9 seconds
calibration operation calibration is performed when the output checker plate is inserted using key operation or external contact
hold / preset operation display and output can be hold or preset by key operation or external contact
calibration curve correction Online correction of input calibration curve, linear to quadratic correction
External Configuration Function Head number, CH.No., calibration, hold, preset possible
Alarm function contact output to HCL terminal when outside the setting range (1 output)
self-diagnostic When detector self-diagnostic is abnormal, contact output (1b) and lamp display
Power supply 100-240V AC 50/60Hz
power consumption 最大15VA
Operating temperature limit 0~50℃
Case Flame Retardant Polycarbonate
Installation Method Panel recessed installation
mass 約0.6kg

Download Documents & Softwares

Catalog

Specification Sheet (PS Sheet)

CAD Data

*Downloading materials is a members-only service.
*If you wish to download, please login as a member and click "Download."

Instruction Manual

*Downloading materials is a members-only service.
*If you wish to download, please login as a member and click "Download."
  • Setting display IRGMEG3 [General]
    ファイルサイズ3MB
    Models
    IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01,IRGMEG
    Document No.
    UX-MIM0219-P6
    Downloads
  • Setting display IRGMEG3 [Communication]
    ファイルサイズ3MB
    Models
    IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01,IRGMEG
    Document No.
    UX-MIM0835-P2
    Downloads
  • Infrared moisture and thickness gauge detector [Communication] IRMA□□□□L□, IRMD□□□□L□
    ファイルサイズ1MB
    Models
    IRGMEG,IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01
    Document No.
    IM-831-P1
    Downloads
  • English)IRGMEG3 Infrared Multi Analyzer Setting
    ファイルサイズ1MB
    Models
    IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01,IRGMEG
    Document No.
    INE-531-P0
    Downloads
  • English)IRGMEG3 IRInfrared Multi Analyzer Setting Display Unit
    ファイルサイズ3MB
    Models
    IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01,IRGMEG
    Document No.
    UX-MIM0219-P6
    Downloads
  • English)IRGMEG3 For Infrared Multi Analyzer Detector unit / Setting display unit
    ファイルサイズ1MB
    Models
    IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01,IRGMEG
    Document No.
    INE-530-P0
    Downloads

Softwares

Application Examples

RelatedInfo

movie

FAQ

If you have any problems with temperature control,
 please feel free to contact us.

Membership
CHINO Web Membership Information

If you register as a CHINO Web member, you will be able to view the product's instruction manual.
If you wish, we will send you the latest information on CHINO products via email.

You must be a login to download.