Explosion explosion proof infrared moisture and thickness meter (Japan Stadnards)

IRMD is a pressure-resistant explosion proof specifications product of the infrared moisture and thickness gauge IM series.
It is a pressure-resistant explosion proof type that can be used in solvent specifications hazardous locations.
It is a moisture ・thickness meter that can measure moisture, organic content, film thickness, coating thickness, color density, etc., using the infrared absorption light that the measurement object has.
Using up to 10 wavelengths in the near-infrared to infrared wavelength range, it can measure a variety of constituent quantities.
In addition, up to 4 constituent simultaneous measurements can be made.

Real-time measurement of various constituent, moisture and thickness gauge using visible and infrared absorption

  • Measurements can be made even in explosion proof areas
  • Absorbance measurement possible for up to 10 wavelengths
  • Up to four constituent can be measured simultaneously
  • A wide range of models to suit the measurement target
  • high-speed and high sensitivity measurement, multi-computation
  • Easy maintenance design with self-diagnostic function

Other features

The compact design saves space, including the main body and explosion proof parts.

Specifications

Model list

Explosion explosion proof

Name Models Type inspection approval number
Mirror type reflective type General purpose moisture IRMD1100 No. TC16974
high moisture type IRMD1200
Trace moisture IRMD1300 No. TC16975
moisture /thickness (color density) IRMD31□□ No. TC16974
Color density (thin film/infrared) IRMD32□□□□ No. TC16975
moisture /Thickness (Near Infrared) IRMD51□□ No. TC16974
moisture /Thickness (thin film/infrared) IRMD52□□ No. TC16975
Thickness/coating amount IRMD7100 No. TC16974
Thin film thickness/coating amount IRMD7200 No. TC16975
Fiber type (reflection/transmission) General purpose moisture, liquid concentration IRMD2100 No. TC16976
high moisture type, liquid concentration IRMD2200
moisture /thickness (color density) IRMD41□□
moisture /Thickness (Near Infrared) IRMD61□□
Thickness/coating amount IRMD8100

*For fiber type, a separate fiber is required.

Specifications

detector

Measurement method infrared absorption type
* moisture and thickness (color density) specifications are based on visible and infrared absorption type
Number of measurement wavelengths Up to 10 wavelengths
Number of constituent measured Up to 4 constituent
* moisture and thickness (color concentration) specifications include up to two constituent in addition to color concentration.
Measurement distance/diameter Mirror type: Measurement distance: 200 to 400 mm
Measurement diameter 50 x 50 mm
Fiber type: Without lens: φ20/15mm to φ40/50mm
With lens: φ20/25mm to φ40/100mm
Output Signal 1. analog signal: 4 to 20mA DC, ±0.2% FS (load resistance 500Ω or less)
②Communication signal: RS-485 (MODBUS)...standard
③Ethernet (LAN)
*③ cannot be used simultaneously with RS-485
Output Update Period Fastest 28ms
* moisture and thickness (color density) specifications are 56ms
display /Settings Data display, setting value display
calibration curve Linear to cubic equations and multiple regression equations
calibration curve correction function available (1st to 2nd order correction)
Number of calibration curve 99 (maximum)
computation Function 2-color, 3-color ratio computation, multiple regression computation,
Color density computation (moisture /Thickness (Color Density) specifications only)
smoothing (delay) 0 to 9.9 seconds, 10 to 99 seconds (can selectable)
calibration calibration possible using checkerboard
self-diagnostic function When self-diagnostic abnormality occurs, contact and communication output
Correction input function External 4 to 20mA DC (1 input) corrects measurement data (sample temperature correction, etc.)
contact Input/Output contact input (Di) preset, data hold, real/smooth switching can be selected by contact input.
contact output (Do) outputs the selected function, either self-diagnostic function (1b) or higher and lower limit alarm (1a)
Operating temperature limit 0 to 45°C (air cooling required for temperatures above 40°C. However, the air must be dry instrument air with a temperature of approximately 30°C or less.)
Power supply 24V DC (supplied from power supply unit IR-WEP. Power supply unit: 100-240V AC, 47-450Hz)
power consumption 約30VA
connection terminal connection (wire feedthrough hardware)
Case Aluminum castings
mass 約10kg
Installation Method Bolt hanging method (four M8 bolts)
explosion proof structure symbol ExdⅡBT5
models Approval Number No. TC16974 (IRMD1110, 1200, 31□□, 51□□, 7100)
No. TC16975 (IRMD1300, 32□□, 52□□, 7200)
No. TC16976 (IRMD2100, 22□□, 41□□, 61□□, 8100)

Explosion explosion proof standard

explosion proof structure symbol ExdⅡBT5
Hazardous characteristic classification pressure-resistant explosion proof models
Applicable gas types Acetone, ammonia, carbon monoxide, ethanol, ethyl methyl ketone, ethylene, vinyl chloride, octane, acetic acid, ethyl acetate, hydrogen cyanide, toluene, butane, propane, hexane, benzene, methanol, methane, gasoline, tetrahydrofuran

Setting display

Models IRGMEG3R IRGMEG3A IRGMEG3S
Input Signal RS-485 (connects to detector), up to 9 units can be connected
analog Output analog signal: 4 to 20mADC, 3 outputs
Output 1/2... load resistance 600Ω or less
Output 3... load resistance 400Ω or less
Communication Output RS-232C RS-422A RS-485
Output scaling Numeric keypad settings (0.1 step)
Output Update Period Communication output: 28 ms x number of detector
display ①Measurement value display, □□□□.□ (decimal point position variable)
②Head number, CH.No., parameter display
Smoothing Smoothing time setting for smoothing (delay): T = 0.1 to 99.9 seconds
calibration operation calibration is performed when the output checker plate is inserted using key operation or external contact
hold / preset operation display and output can be hold or preset by key operation or external contact
calibration curve correction Online correction of input calibration curve, linear to quadratic correction
External Configuration Function Head number, CH.No., calibration, hold, preset possible
Alarm function contact output to HCL terminal when outside the setting range (1 output)
self-diagnostic When detector self-diagnostic is abnormal, contact output (1b) and lamp display
Power supply 100-240V AC 50/60Hz
power consumption 最大15VA
Operating temperature limit 0~50℃
Case Flame Retardant Polycarbonate
Installation Method Panel recessed installation
mass 約0.6kg

Fiber Section

>Measurement distance/diameter Without lens φ20/15mm~φ40/50mm
With lens φ20/25mm~φ40/100mm
Fiber Length Standard 1.5m, maximum 5m (reflective fiber)
Standard 2m, Maximum 10m (through-beam fiber)
Fiber Protection Stainless steel coil (reflective fiber optic)
Teflon surface treatment (through-type fiber)
Minimum bending radius R100mm (reflective fiber)
R250mm (through-beam fiber)
Operating temperature range *1 0~150℃
Purge air function Without lens: 5 to 20 NL/min
With lens...No purge function
accessories Horizontal mounting holder, flange holder
*1 specifications differ between reflective and through-beam fibers.

Download Documents & Softwares

Catalog

Specification Sheet (PS Sheet)

  • IM Series Explosion explosion proof Infrared moisture and Thickness Gauge (Mirror Type, Fiber Type)
    ファイルサイズ1MB
    Document No.
    PP-106-5

CAD Data

*Downloading materials is a members-only service.
*If you wish to download, please login as a member and click "Download."

Instruction Manual

*Downloading materials is a members-only service.
*If you wish to download, please login as a member and click "Download."
  • Setting display IRGMEG3 [General]
    ファイルサイズ3MB
    Models
    IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01,IRGMEG
    Document No.
    UX-MIM0219-P6
    Downloads
  • Setting display IRGMEG3 [Communication]
    ファイルサイズ3MB
    Models
    IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01,IRGMEG
    Document No.
    UX-MIM0835-P2
    Downloads
  • Infrared moisture and thickness gauge detector [Communication] IRMA□□□□L□, IRMD□□□□L□
    ファイルサイズ1MB
    Models
    IRGMEG,IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01
    Document No.
    IM-831-P1
    Downloads
  • English)IRGMEG3 Infrared Multi Analyzer Setting
    ファイルサイズ1MB
    Models
    IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01,IRGMEG
    Document No.
    INE-531-P0
    Downloads
  • English)IRGMEG3 IRInfrared Multi Analyzer Setting Display Unit
    ファイルサイズ3MB
    Models
    IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01,IRGMEG
    Document No.
    UX-MIM0219-P6
    Downloads
  • English)IRGMEG3 For Infrared Multi Analyzer Detector unit / Setting display unit
    ファイルサイズ1MB
    Models
    IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01,IRGMEG
    Document No.
    INE-530-P0
    Downloads
  • English)IRMA Infrared multiple-constituents analyzer detector unit LAN communication
    ファイルサイズ1MB
    Models
    IRMA3,IRMA4,IRMA7800S,IRMA_Fiber,IRMA_Mirror,IRMD_Fiber,IRMD_Mirror,IRMT01
    Document No.
    INE-488-P1
    Downloads

Softwares

Application Examples

RelatedInfo

movie

FAQ

If you have any problems with temperature control,
 please feel free to contact us.

Membership
CHINO Web Membership Information

If you register as a CHINO Web member, you will be able to view the product's instruction manual.
If you wish, we will send you the latest information on CHINO products via email.

You must be a login to download.