The IM series is moisture /thickness meter that can measure moisture, organic matter, film thickness, coating thickness, color density, etc. by utilizing the infrared absorption of the object being measured.
Up to 10 wavelengths can be used in the wavelength range from near infrared to infrared, allowing the amount of a variety of constituent to be measured.
In addition, up to four constituent can be measured simultaneously.
The compact design saves space.
Name | Models | |
---|---|---|
Mirror type reflective type | General purpose moisture | IRMA1100□□ |
High moisture type | IRMA1200□□ | |
Micro moisture | IRMA1300□□ | |
Moisture /thickness (color density) | IRMA31□□□□ ** | |
Moisture /Thickness (Near Infrared) | IRMA51□□□□ | |
Moisture /Thickness (thin film/infrared) | IRMA52□□□□ | |
Thickness/coating amount | IRMA7100□□ | |
Thin film thickness/coating amount | IRMA7200□□ | |
Fiber type (reflection/transmission) | General purpose moisture, liquid concentration | IRMA2100□□ |
High moisture type, liquid concentration | IRMA2200□□ | |
Moisture /thickness (color density) | IRMA41□□□□ ** | |
Moisture /Thickness (Near Infrared) | IRMA61□□□□ | |
Thickness/coating amount | IRMA8100□□ |
*For fiber type, a separate fiber is required. ** (Japan Only)
Measurement method | Infrared absorption type * moisture and thickness (color density) specifications are based on visible and infrared absorption type |
|
---|---|---|
Number of measurement wavelengths | Up to 10 wavelengths | |
Number of constituent measured | Up to 4 constituent * moisture and thickness (color concentration) specifications include up to two constituent in addition to color concentration. |
|
Measurement distance/diameter | Mirror type: | Measurement distance: 200 to 400 mm Measurement diameter 50 x 50 mm□ |
Fiber type: | Without lens: φ20/15mm to φ40/50mm With lens: φ20/25mm to φ40/100mm |
|
Output Signal | 1. Analog signal: 4 to 20mA DC, ±0.2% FS (load resistance 500Ω or less) ②Communication signal: RS-485 (MODBUS)...standard ③Ethernet (LAN) *③ Cannot be used simultaneously with RS-485 |
|
Output Update Period | Fastest 28ms * moisture and thickness (color density) specifications are 56ms |
|
Display /Settings | Data display, setting value display, various settings can be keyed in | |
Calibration curve | Linear to cubic equations and multiple regression equations calibration curve correction function available (1st to 2nd order correction) |
|
Number of calibration curve | 99 (maximum) | |
Computation Function | 2-color, 3-color ratio computation, multiple regression computation Color density computation (moisture /thickness (color density meter) only) |
|
Smoothing (delay) | 0 to 9.9 seconds, 10 to 99 seconds (can selectable) | |
Calibration | Calibration possible using checkerboard | |
Self-diagnostic function | When self-diagnostic abnormality occurs, contact and communication output | |
Correction input function | External 4 to 20mA DC (1 input) corrects measurement data (sample temperature correction, etc.) | |
Contact Input/Output | Contact input (Di) preset, data hold, real/smooth switching can be selected by contact input. contact output (Do) outputs the selected function, either self-diagnostic function (1b) or higher and lower limit alarm (1a) |
|
Operating temperature limit | 0 to 50°C (air cooling required for temperatures above 45°C. However, the air must be dry instrument air with a temperature of approximately 30°C or less.) | |
Power supply | 24V DC (supplied from power supply unit IR-WEP. Power supply unit: 100-240V AC, 47-450Hz) | |
Power consumption | 約30VA | |
Connection | terminal connection (wire feedthrough hardware) | |
Case | Aluminum castings | |
Weight | 約4.3kg | |
Installation Method | Bolt hanging method (four M8 bolts) | |
CE marking | CE marking compliant EMC directive EN61326+A |
Models | IRGMEG3R | IRGMEG3A | IRGMEG3S |
---|---|---|---|
Input Signal | RS-485 (connects to detector), up to 9 units can be connected | ||
Analog Output | analog signal: 4 to 20mADC, 3 outputs Output 1/2... load resistance 600Ω or less Output 3... load resistance 400Ω or less | ||
Communication Output | RS-232C | RS-422A | RS-485 |
Output scaling | Numeric keypad settings (0.1 step) | ||
Output Update Period | Communication output: 28 ms x number of detector | ||
Display | ①Measurement value display, □□□□.□ (decimal point position variable) ②Head number, CH.No., parameter display |
||
Smoothing | Smoothing time setting for smoothing (delay): T = 0.1 to 99.9 seconds | ||
Calibration operation | calibration is performed when the output checker plate is inserted using key operation or external contact | ||
Hold / preset operation | display and output can be hold or preset by key operation or external contact | ||
Calibration curve correction | Online correction of input calibration curve, linear to quadratic correction | ||
External Configuration Function | Head number, CH.No., calibration, hold, preset possible | ||
Alarm function | contact output to HCL terminal when outside the setting range (1 output) | ||
Self-diagnostic | When detector self-diagnostic is abnormal, contact output (1b) and lamp display | ||
Power supply | 100-240V AC 50/60Hz | ||
Power consumption | 最大15VA | ||
Operating temperature limit | 0~50℃ | ||
Case | Flame Retardant Polycarbonate | ||
Installation Method | Panel recessed installation | ||
Weight | 約0.6kg |
Measurement distance/diameter | Without lens | φ20/15mm~φ40/50mm |
---|---|---|
With lens | φ20/25mm~φ40/100mm | |
Fiber Length | Standard 1.5m, maximum 5m (reflective fiber) Standard 2m, Maximum 10m (through-beam fiber) |
|
Fiber Protection | Stainless steel coil (reflective fiber optic) Teflon surface treatment (through-type fiber) |
|
Minimum bending radius | R100mm (reflective fiber) R250mm (through-beam fiber) |
|
Operating temperature range *1 | 0~150℃ | |
Purge air function | Without lens: 5 to 20 NL/min With lens...No purge function |
|
Accessories | Horizontal mounting holder, flange holder |
The IM series is moisture /thickness meter that can measure moisture, organic matter, film thickness, coating thickness, etc. by utilizing the infrared absorption of the object being measured.
IRMA/IRMD/IRMT01
The IM series is moisture /thickness meter that can measure moisture, organic matter, film thickness, coating thickness, color density, etc. by utilizing the infrared absorption of the object being measured.
IRMA1000/2000 series/IRMD1000/2000 series
The IM series is moisture /thickness meter that can measure moisture, organic matter, film thickness, coating thickness, etc. by utilizing the infrared absorption of the object being measured.
IRMA7000/8000 series/IRMD7000/8000 series
The IM series is moisture /thickness meter that can measure moisture, organic matter, film thickness, coating thickness, etc. by utilizing the infrared absorption of the object being measured.
IRMA2000, 4000, 6000 series/IRMD2000, 4000, 6000 series
The visible/infrared hybrid IM series is a high-speed response constituent meter that can simultaneously measure color concentration (255 levels of color difference), turbidity (transparency), moisture, thickness, and other constituent online and non-contact.
IRMA3/IRMA4/IRMD3/IRMD4
If you register as a CHINO Web member, you will be able to view the product's instruction manual.
If you wish, we will send you the latest information on CHINO products via email.
You must be a login to download.