The TP-U series is a body surface temperature measurement unit equipped with a 2000-pixel thermal image sensor.
Models | Specifications |
---|---|
TP-U0260EC | With manual temperature compensation |
TP-U0260ES | With automatic temperature compensation function |
Manual correction specifications TP-U0260EC |
Automatic correction specifications TP-U0260ES |
|
Temperature Measurement Range | 20℃~40℃ | |
Measurement temperature compensation | Manually set the correction value from the computer | Automatic calculation based on the measurement data of the set number of samples |
Accuracy ratings | ±0.5℃ when measured at 37℃ (ambient temperature 25℃) | |
Detecting element | Thermopile element 2000 pixel | |
Wavelength | Central wavelength 10μm | |
Measurement field of view angle | 60°×60° | |
radius resolution | 21.8mrad | |
Frame rate | 1Hz | |
Interface | 10BASE-T/100BASE-TX | |
focus | Fixed Focus | |
Number of contact outputs | 2 points (non-voltage contact output) | |
Data Storage | Measurement data and thermal images (JPEG) are automatically saved to a PC | |
Power supply | 100-240V AC (universal power supply) | |
Operating temperature limit | 0℃~40℃ | |
Operating humidity range | 10 to 80% RH (but no dew condensation) | |
Weight | 約0.7kg |