Add finely crushed ice and water (preferably pure water) to a thermos or other container with good thermal insulation properties. After making sure that the ice is not melting (stable state), measure the surface temperature of the ice and water. (Be careful not to cause any field of lack of view.) At this time, it is necessary to set emissivity of radiation thermometer to 1.0. The surface temperature of the ice and water in this state will be approximately 0°C, so a simple scale check can be made by checking indicated value on radiation thermometer. (However, please check the reading accuracy of radiation thermometer before proceeding.) (You need to put crushed ice in a container and dig a groove in it to increase the apparent emissivity.) |
If you register as a CHINO Web member, you will be able to view the product's instruction manual.
If you wish, we will send you the latest information on CHINO products via email.
You must be a login to download.